CT-Less Attenuation Correction Using Multiview Ensemble Conditional Diffusion Model on High-Resolution Uncorrected PET Images
2510.24805v1
q-bio.QM, cs.AI, cs.CV
2025-11-01
Авторы:
Alexandre St-Georges, Gabriel Richard, Maxime Toussaint, Christian Thibaudeau, Etienne Auger, Étienne Croteau, Stephen Cunnane, Roger Lecomte, Jean-Baptiste Michaud
Abstract
Accurate quantification in positron emission tomography (PET) is essential
for accurate diagnostic results and effective treatment tracking. A major issue
encountered in PET imaging is attenuation. Attenuation refers to the diminution
of photon detected as they traverse biological tissues before reaching
detectors. When such corrections are absent or inadequate, this signal
degradation can introduce inaccurate quantification, making it difficult to
differentiate benign from malignant conditions, and can potentially lead to
misdiagnosis. Typically, this correction is done with co-computed Computed
Tomography (CT) imaging to obtain structural data for calculating photon
attenuation across the body. However, this methodology subjects patients to
extra ionizing radiation exposure, suffers from potential spatial
misregistration between PET/CT imaging sequences, and demands costly equipment
infrastructure. Emerging advances in neural network architectures present an
alternative approach via synthetic CT image synthesis. Our investigation
reveals that Conditional Denoising Diffusion Probabilistic Models (DDPMs) can
generate high quality CT images from non attenuation corrected PET images in
order to correct attenuation. By utilizing all three orthogonal views from
non-attenuation-corrected PET images, the DDPM approach combined with ensemble
voting generates higher quality pseudo-CT images with reduced artifacts and
improved slice-to-slice consistency. Results from a study of 159 head scans
acquired with the Siemens Biograph Vision PET/CT scanner demonstrate both
qualitative and quantitative improvements in pseudo-CT generation. The method
achieved a mean absolute error of 32 $\pm$ 10.4 HU on the CT images and an
average error of (1.48 $\pm$ 0.68)\% across all regions of interest when
comparing PET images reconstructed using the attenuation map of the generated
pseudo-CT versus the true CT.