High-Resolution Magnetic Particle Imaging System Matrix Recovery Using a Vision Transformer with Residual Feature Network
2511.02212v1
physics.med-ph, cs.CV, eess.IV
2025-11-06
Авторы:
Abuobaida M. Khair, Wenjing Jiang, Yousuf Babiker M. Osman, Wenjun Xia, Xiaopeng Ma
Abstract
This study presents a hybrid deep learning framework, the Vision Transformer
with Residual Feature Network (VRF-Net), for recovering high-resolution system
matrices in Magnetic Particle Imaging (MPI). MPI resolution often suffers from
downsampling and coil sensitivity variations. VRF-Net addresses these
challenges by combining transformer-based global attention with residual
convolutional refinement, enabling recovery of both large-scale structures and
fine details. To reflect realistic MPI conditions, the system matrix is
degraded using a dual-stage downsampling strategy. Training employed
paired-image super-resolution on the public Open MPI dataset and a simulated
dataset incorporating variable coil sensitivity profiles. For system matrix
recovery on the Open MPI dataset, VRF-Net achieved nRMSE = 0.403, pSNR = 39.08
dB, and SSIM = 0.835 at 2x scaling, and maintained strong performance even at
challenging scale 8x (pSNR = 31.06 dB, SSIM = 0.717). For the simulated
dataset, VRF-Net achieved nRMSE = 4.44, pSNR = 28.52 dB, and SSIM = 0.771 at 2x
scaling, with stable performance at higher scales. On average, it reduced nRMSE
by 88.2%, increased pSNR by 44.7%, and improved SSIM by 34.3% over
interpolation and CNN-based methods. In image reconstruction of Open MPI
phantoms, VRF-Net further reduced reconstruction error to nRMSE = 1.79 at 2x
scaling, while preserving structural fidelity (pSNR = 41.58 dB, SSIM = 0.960),
outperforming existing methods. These findings demonstrate that VRF-Net enables
sharper, artifact-free system matrix recovery and robust image reconstruction
across multiple scales, offering a promising direction for future in vivo
applications.